Altana Haze-Gard Plus Operating Instructions Uživatelský manuál Strana 32

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 180
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 31
Fig. 2
Illumination
Detector
Sphere Exit
Ring Sensor
Specimen
Sphere Entrance
Measurement Principle
A beam strikes the specimen and
enters an integrating sphere
(Fig. 2). The light, uniformly
distributed by the matte white
coating on the sphere wall, is
measured by a detector. Total
transmittance is measured with the
sphere outlet closed, and haze is
measured with the outlet open.
A ring sensor in the outlet opening
measures clarity. The entire
measurement sequence is controlled
automatically.
Standards
ASTM D 1003: Standard Test Method
for Haze and Luminous Transmittance
of Transparent Plastics
ASTM D 1044: Standard Test Method
for Resistance of Transparent Plastics
to Surface Abrasion
Zobrazit stránku 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 179 180

Komentáře k této Příručce

Žádné komentáře