Altana Haze-Gard Plus Operating Instructions Uživatelský manuál Strana 22

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 180
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 21
delete all
delete last
statistic
on/off
result
on/off
delete all
delete last
Change Cal Value
Input Trans Std
Input Clarity Std
Save Std Values
Escape: SET UP
T Std = 100
1000
Confirm: OPERATE
Escape: SET UP
Initialize Instr.
Old New
T Std 100 100
C Std 75.5 75.5
Save Data: OPERATE
Escape: SET UP
Select the
standard you
want to change,
and confirm with
<operate>.
In the four-digit display, the last
number indicates the number of
digits after the decimal point.
Select the
decimal
place to
be
modified.
Change
the value
and
confirm
with
<operate>.
Pressing the <operate> button
saves the new setting for the stan-
dard value and returns you to the
"Change Calibration Value" subme-
nu. Pressing the <set up> button
rejects the change and retains the
old standard value.
These new values remain valid only
until the unit is switched off. When it
is switched back on, the old values
are restored.
The "Save Standard Values"
function serves to save the reset
standard values permanently, so
that they are retained even after
the instrument is switched off. If a
change has been confirmed with
<operate>, a recalibration is neces-
sary after exiting from the "Change
Calibration Value" submenu.
Zobrazit stránku 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 179 180

Komentáře k této Příručce

Žádné komentáře