Altana Spectro-Guide Sphere Gloss Manual Uživatelský manuál Strana 45

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 74
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 44
44
Options
10.5 Calibration
Chk Green
From the “Options” menu, use the
button to move the cursor to the “Calibr.” option.
The submenu opens when you press the Operate
button. The “chk green” option is preselected.
Take the green test standard and open the lid.
Position the instrument so that the display points in
the direction of the open lid and the front edge of the
instrument is parallel to the edge of the lid. Press
the Operate button.
At this point, you will be prompted to measure the
green test standard.
You will then see a message indicating whether the
calibration is still valid. If the message
Please calibrate!
appears, the calibration must be performed with the
black and white standards (see Chapter 6 –
Calibrate).
std
File Display
Color sys
Index
Illum/Obs
Setup Calibrate
Configure
f
f
f
f
f
Option
Calibr. Chk Green
f
Gloss
f
File Display
Measure Green Std
Press Operate Button
Option
File Display
Calibration OK
Option
Zobrazit stránku 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 73 74

Komentáře k této Příručce

Žádné komentáře