9.1 Number of Measurements
This function allows selection of the numbers of
readingstobetakenpersample.Ifthenumberis
greater than 1, the measurements are statistically
evaluated. Turning the scroll wheel adjusts the
number while pressing it shifts the decimal place
one further over. Pressing operate completes the
processandaconrmationdisplayappears.
9.2 Scanlength
The setting options for Scanlength are shown in
the neighboring illustration. The current setting is
identiedbyablackdotattheendoftheline.
Move the cursor to the desired setting and press
the operate button. The instrument returns to the
previous menu.
Inthesetting0cm,youcanperform
measurements without scan motion. However, you
can only determine the scales du, Wa, Wb and
DOIhere.
Note:
Using a short scanlength will result in strong
statistical skewing of measurement values for large
wavelengths. Thus, when using the Wd scale with
a scanlength of 5 cm, at least 3 measurements
should be made per checkzone to ensure
representative values.
Conguration
Scanlength
0cm
5cm
10 cm
20 cm
l