Do not attempt to repair the unit yourself. If your measuring unit should
experience a malfunction, our costumer service will be glad to help you
as quickly as possible.
Fluctuations of
the measured
values
Was the same spe-
cimen position used
for all the measure-
ments?
No: Check how high
the deviations are on
the specimen itself.
Error Messages 15
Yes: If applicable,
check the calibration
using the auxiliary
standard.
Calibration incorrect:
Recalibrate, possibly
clean the standard
(chap. 7).
Calibration correct:
Is the specimen sur-
face completely
smooth and does the
measuring unit make
good contact with the
specimen?
No: In this case, very
large deviations are
possible (chap. 10).
Yes: The unit is
defective.
Please contact our
customer service.