Altana LCM IV Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Altana LCM IV. ALTANA LCM IV User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 113
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků

Shrnutí obsahu

Strany 1 - LCS IV / LCM IV

Additives & InstrumentsA member ofLCS IV / LCM IVMeasure what you see.Manual

Strany 2 - Table of Contents

11General InformationThe LCS IV contains the following programs and operating modes in addition to the color measurement: single wavelength mode, mult

Strany 3

101Maintenance1. Switch off instrument.2. Remove the power cable from the instrument.3. Open the cuvette compartment cover and remove cuvette adapter

Strany 4

102Maintenance

Strany 5

103Section 8 TroubleshootingError displayed Definition ResolutionAn error occurred when uploading the instrument data.Restart the process or contact t

Strany 6 - Section 1 Specifications

104TroubleshootingFaultTest program stopped!Please check lampClose the lid.Error [xx]Test program stops when the instrument is started Check the lamp

Strany 7 - Specifications

105TroubleshootingSystem check incorrect! Measurement of air values failedSwitch the instrument off and then back on again. If the system test is unsu

Strany 9 - 2.2 Overview of product

107Section 9 Service and Certification DeutschlandBYK-Gardner GmbHLausitzer Strasse 882538 GeretsriedDeutschlandTel.: +49-8171-3493-0Fax: +49-8171-34

Strany 10 - General Information

108Service and Certification

Strany 11

109IndexAAbsorbance Formula ... 74Acid wash color ...

Strany 12 - Section 3 Installation

110IndexOverview of Product/Function ... 10PPassword ... 27, 3

Strany 14 - 3.4 Power connections

(&'HFODUDWLRQRI&RQIRUPLW\:H %<.*DUGQHU*PE+/DXVLW]HU6WUDVVH*HUHWVULHGKHUHZLWKGHFODUHWKDWWKHSURGXFWV7\SH /&6,9

Strany 15 - 3.5 Interfaces

'«FODUDWLRQGHFRQIRUPLW«&(1RXVOಬHQWUHSULVH %<.*DUGQHU*PE+/DXVLW]HU6WUDVVH*HUHWVULHGG«FODURQVSDUODSU«VHQWHTXHOHVSURG

Strany 16 - Installation

'HFODUDFLµQGHFRQIRUPLGDG&(1RVRWURVODVRFLHGDG %<.*DUGQHU*PE+/DXVLW]HU6WUDVVH*HUHWVULHGGHFODUDPRVSRUODSUHVHQWHTXHORVS

Strany 17

235 022 649 E 1211

Strany 18 - Section 4 Start Up

13Section 3 Installation3.1 Unpacking the instrumentThe following components are supplied as standard with the LCS IV / LCM IV:• LCS IV / LCM IV spect

Strany 19 - 4.5 Power off the instrument

14Installation3.3 Front and back viewFigure 1 Front view 1 USB port type A 3 Touch screen2 Cuvette compartment cover 4 On/off switch

Strany 20 - Section 5 Standard programs

15InstallationFigure 2 Back view 3.4 Power connections1. Connect the power cable to the benchtop power supply. 2. Plug the benchtop power supply ca

Strany 21 - 5.1.3 Main menu

16Installation3.5 InterfacesThe instrument has three USB ports and one Ethernet port as standard. They are located on the front and rear of the instr

Strany 22 - 5.2 Instrument Setup

17InstallationCell compartment (2) for:The following cell types can be used in cell compartment (2).• Without cuvette adapter Z in the cuvette compart

Strany 23 - 5.2.2 Sample ID

18InstallationFigure 3 Cuvette compartments and cuvette adapter Z 1 Cuvette compartment (1) for round cuvettes 2 Cuvette compartment (2) for square

Strany 24

19Section 4 Start Up4.1 Switch on the instrument, startup process1. Connect the power cable to the mains outlet.2. Switch on the instrument by pressin

Strany 25

20Start UpThe Main Menu is displayed when diagnostics are completed.Note: section 5.1.3, page 22 contains a detailed description of the main menu.4.4

Strany 26 - 5.2.3 Security settings

3Table of ContentsSection 1 Specifications ...

Strany 27

21Section 5 Standard programs5.1 Overview5.1.1 Tips for using the touch screenThe whole screen responds to touch. To choose an option, tap with a fing

Strany 28

22Standard programs5.1.3 Main menuA variety of modes may be selected from the Main Menu. The following table briefly describes each menu option.There

Strany 29 - 5.2.4 Date and time

23Standard programs5.2 Instrument Setup1. In the main menu, select the menu option Instrument setup.A selection of functions appears in order to confi

Strany 30

24Standard programs10. With Back the device returns to the menu "Instrument Setup".Press Logout to log out of an active operator ID.Press Lo

Strany 31 - 5.2.6 PC and printer

25Standard programs6. To number the sample IDs sequentially (e. g. Inflow (01) etc.), press Add Number. • Use the arrow keys to specify the first numb

Strany 32

26Standard programs4. The sample ID can have the current date and time, a sequential number and a colour allocated to it.Select the required options o

Strany 33

27Standard programs5. Select the desired file with OK.6. Transfer the displayed sample ID list with DONE.5.2.3 Security settingsThe "Security&quo

Strany 34

28Standard programs5. Select the required function and press Setup.6. Select the required security level (two keys, one key or off) and confirm using

Strany 35

29Standard programs4. Press Operator Password.5. Enter the operator password and confirm with OK.6. Press Security Level: <Off>.7. Enter the sec

Strany 36

30Standard programsThe operator ID is displayed with the selected security level. 10. Activate the selected operator ID by pressing Login.11. Enter th

Strany 37 - 5.3.1 Color values

4Table of Contents5.2 Instrument Setup...

Strany 38

31Standard programs2. Press Date Format for the date.3. Press OK to confirm the entry.4. Press Time Format for the time.5. Press OK to confirm the ent

Strany 39

32Standard programs5.2.5 Sound settings 1. Press Sound Settings in "Instrument Setup".The following options will be displayed:• All: Activat

Strany 40

33Standard programs1. Press PC & Printer in "Instrument Setup".A list with information about the connections opens.5.2.6.1 Printer setup

Strany 41

34Standard programs6. Press Paper to select the paper size.Select between:• Letter,• Monarch,• Executive,• A4.7. Press OK to confirm the entry.Note: P

Strany 42 - 5.3.3 Data log (only LCS IV)

35Standard programs5.2.6.4 Network setupThe instrument supports communication in local networks. For the setup of an Ethernet communication in network

Strany 43

36Standard programsIn conjunction with the IP address of an instrument, the subnet mask establishes which IP addresses are located within the local ne

Strany 44

37Standard programsThe target site can be defined by IP address or a server name. 12. Press IP-Address and enter the IP address or select Server Name

Strany 45

38Standard programs5.2.7 Power Management1. Press Power Management in "Instrument setup".2. Press Sleep Timer and then OK.3. Select the requ

Strany 46

39Standard programs5.3.1.1 Automatic/manual color value storageThe data storage function indicates whether data is to be stored automatically or manua

Strany 47

40Standard programs3. Press Filter: On/Off. The function Filter Settings is used to search for specific items.4. Activate On. The data can now be filt

Strany 48

5Table of Contents5.5 Color measurement ...

Strany 49 - 5.3.5 Trends

41Standard programs9. Press Control Limit.10. Highlight On to change the Upper Control Limit and the Lower Control Limit settings.11. The measured poi

Strany 50

42Standard programs4. Press Options and then the PC & Printer icon.5. Select the data to be sent.The following options are available:• Single poin

Strany 51

43Standard programs2. Press AQA Log.A listing of the stored data is displayed.3. Press Filter: On/Off. 4. Activate On. 5. Select a required parameter.

Strany 52 - 5.4 Take and prepare samples

44Standard programs5.3.3.1 Automatic/manual saving of single and multi-wavelength measurementsThe data log function indicates whether data is to be st

Strany 53 - 5.5 Color measurement

45Standard programs3. Press Filter: On/Off. The function Filter Settings is used to search for specific items.4. Activate On. The data can now be filt

Strany 54

46Standard programs9. Press Control Limit.10. Highlight On to change the Upper Control Limit and the Lower Control Limit settings.11. The measured poi

Strany 55

47Standard programs4. Press Options and then the PC & Printer icon.5. Select the data to be sent.The following options are available:• Single poin

Strany 56

48Standard programs5.3.4.1 Data storage from wavelength scan or time course 1. Press the Folder icon in the "Options" menu after a reading i

Strany 57

49Standard programs4. Press Table to display details.Note: Press View List to return to the "Recall Data" list.5.3.4.3 Send data from wavele

Strany 58

50Standard programsOption 2: 1. Press Wavelength Scan or Time Course and then Options > More > Send Data to send the data to a USB memory stick

Strany 59

6Table of ContentsSection 6 Advanced Operations ...

Strany 60

51Standard programs1. Press Recall Data > Trends.2. A list of all trend analyses is displayed.3. Press Options and New to create a new trend.Edit

Strany 61

52Standard programs5. Use control limits to define the upper and lower limits that are displayed in the graphic.6. Select the required trend and confi

Strany 62

53Standard programs9. Press Add Data to include additional measured values in the trend.5.4 Take and prepare samplesTake a representative sample from

Strany 63

54Standard programs• If air bubbles are entrapped, remove them by heat, vacuum, ultrasonic treatment or other suitable means. • Clean the outside of t

Strany 64

55Standard programs5.5.1 Take a color measurement1. Press Color Measurement.2. Insert a cuvette/sample cell with distilled water to calibrate.Note: Al

Strany 65

56Standard programsFigure 4 Touch-sensitive areas in measurement mode 5.5.1.2 Parameter setup optionsPress Options to set up the parameter.1 Open Se

Strany 66

57Standard programs5.5.1.3 Change the color scale after a measurement1. Press the touch-sensitive area, e. g. Color Value Gardner. 2. A list of all co

Strany 67 - Table 6 contains

58Standard programsThe current result of the actual reading and all further measurements will be displayed in the selected color scale.Alternative1. T

Strany 68

59Standard programscuvettes. Hazen values below 50 should be measured in 50 mm cuvettes. Hazen values below 10 should be measured only with the 50 mm

Strany 69

60Standard programsThere are two ways of color matching of a sample with the Ph.Eur.system.system:• Automatic: The color of the solution is compared t

Strany 70 - 5.7 Photometry (LCS IV only)

7Section 1 SpecificationsThese are subject to change without notice!Performance specifications LCS IV LCM IVDisplay modeColor measurement, color diffe

Strany 71

61Standard programs3. Press Auto.4. Press Graph.The color matching result is now displayed graphically. The upper value shows the option used for the

Strany 72

62Standard programsThe CIE L*a*b* values depend on the path length of the cuvette/sample cell. Therefore the scale of the a*, b* axis in the graph dep

Strany 73

63Standard programsGraph for the red scale R.If the CIE L*a*b* value of the sample does not correspond to any reference solution in a selected scale,

Strany 74

64Standard programsThe CIE L*a*b* values depend on the path length and are always related to the type of cuvette/sample cell used.1. Press US Pharmaco

Strany 75

65Standard programs5.5.7.5 Determination of the AOCS Cc 13 e (Lovibond®) scaleThe calculation of the Lovibond color values is derived from the methods

Strany 76

66Standard programs5.5.7.9 Determination of the Acid wash colorASTM D848 - 09 is the standard test method for the acid wash color in the "aromati

Strany 77

67Standard programs5.6 Measurement of color differences (LCS IV only)The mode for color difference measurement is used to determine a quantitative col

Strany 78

68Standard programs7. Press Option for Parameter Setup. Table 6 contains additional information on the program options. Table 6 Options for color dif

Strany 79

69Standard programs5.6.1.1 View graph/table/values1. Press Options > View Graph to display the transmittance graph for the sample (black line) and

Strany 80

70Standard programs5.6.2 Take a color difference measurement with stored reference values1. Press Color Difference Measurement from the Main Menu. 2.

Strany 81

8SpecificationsInterfacesUse only shielded cable with maximum length of 3 m:2× USB type A1× USB type BUse only shielded cable (for example STP, FTP, S

Strany 82

71Standard programs5.7 Photometry (LCS IV only)In the photometry mode, photometric measurements can be carried out as: • Single wavelength measurement

Strany 83

72Standard programsConcentration Factor:1. Press Concentration Factor: Off in the "Options" menu. 2. Press On to activate this function.3.

Strany 84

73Standard programs5.7.1.2 Take single wavelength readings (single reading)1. Insert the zero solution cell into the cell compartment.2. Press Zero.No

Strany 85

74Standard programsAbsorbance Formula1. Press Absorbance Formula. 2. The formula selected in the top key determines the number of wavelength and coeff

Strany 86 - Section 6 Advanced Operations

75Standard programsK 1 refers to the coefficient at wavelength 1K 2 refers to the coefficient at wavelength 2, etc. If a subtraction has to be complet

Strany 87 - 6.1.3 Optical Checks

76Standard programs5.7.2.2 Complete a reading in the Multi Wavelength mode1. Insert the zero solution cell into the cell compartment.2. Press Zero. No

Strany 88

77Standard programsTime & Interval:1. Press the "Options" menu on the Time & Interval menu option.2. Enter the total time and measur

Strany 89

78Standard programs5.7.3.2 Time course scan readingAfter all scan parameters have been selected, the instrument must then be zeroed. Afterward the sam

Strany 90 - 6.1.5 Instrument backup

79Standard programs5.7.4 Wavelength Scan mode – recording of absorbance and transmission spectrumsIn the Wavelength Scan mode, the absorbance of the l

Strany 91 - 6.1.7 Service time

80Standard programs5.7.4.1 Set up the wavelength scanIn the main menu, press the Wavelength Scan menu option. Press Options for Parameter Setup.Table

Strany 92 - 6.2 Toolbar

9Section 2 General Information2.1 Safety informationRead through the entire user manual carefully before you unpack the device, set up and put into op

Strany 93 - 6.2.4 AQA

81Standard programsSet the wavelength range1. Press the Okey in the "Options" menu to select the wavelength range and the wavelength step.2

Strany 94 - 6.2.5 Trends

82Standard programsIntegralThe Integral applies to the whole wavelength range of the scan.1. Press Integral: Off in the "Options" menu.2. Hi

Strany 95

83Standard programs3. Press Zero. During the scan of the base line, "Zero calibration" is shown in the screen.4. Insert the prepared analysi

Strany 96 - Section 7 Maintenance

84Standard programs2. The reference curve is displayed in orange. The absorbance or transmittance value and the associated wavelength are highlighted

Strany 97 - 7.2 Lamp replacement

85Standard programs4. The reference graph is displayed in orange. The absorbance or transmittance value and the associated wavelength are displayed in

Strany 98

86Standard programs

Strany 99

87Section 6 Advanced Operations6.1 System Checks1. Press System Checks in the Main Menu.The "System Check" menu contains instrument informat

Strany 100 - Maintenance

88Advanced Operations1. Tap on Instrument Update in the "System Checks" menu.2. Plug the USB storage device into the USB port (USB A) of the

Strany 101

89Advanced Operations7. Remove any present cuvettes from the cuvette compartment.8. Close the cuvette compartment cover.9. Press Start.10. Follow the

Strany 102 - Section 8 Troubleshooting

90Advanced Operations2. Select Options > New to define a new standard. 3. Enter the name of the standard solution and confirm with OK.4. Enter the

Strany 103 - Troubleshooting

10General Information2.1.2 Safety around source lampsThe source lamp is operated at high temperatures.To prevent electric shock, disconnect the instru

Strany 104

91Advanced Operations7. Select Options for further editing.Select New to define a new standard.Select Edit to change the settings.Select Delete to del

Strany 105

92Advanced OperationsRestore backup data:Note: All current data will be overwritten when restoring the Backup file!1. Press Instrument Backup in the &

Strany 106

93Advanced Operations6. Activate the options field under Next service.7. Select a time interval for the next service.8. Press OK to confirm the select

Strany 107 - Service and Certification

94Advanced Operations6.2.2 Sample IDUse this function to enter up to 100 different sample designations (each with max. 20 characters) into the instrum

Strany 108

95Advanced Operations6.2.5 TrendsUse this function to call up a load curve of the stored measurement values for each parameter or sample location. The

Strany 109

96Advanced Operations

Strany 110 - (*.RQIRUPLW¦WVHUNO¦UXQJ

97Section 7 Maintenance7.1 Cleaning requirements7.1.1 Spectrophotometer• Only clean the housing, the cell compartments and all accessories with a soft

Strany 111

98Maintenance7.1.3 Cuvettes/cells1. After use, clean the glass cells with cleaning agents.2. Afterwards, rinse the cuvettes/sample cells several times

Strany 112

99Maintenance6. Using a cross-head screwdriver, remove the two screws from the lamp holder.7. Lift out the lamp holder upwards.8. Turn the lamp holder

Strany 113 - 235 022 649 E 1211

100MaintenanceTo fit a new lamp carry out the procedure in reverse.1. Connect a new lamp to the connector.2. Install the lamp with the connector in th

Příbuzné modely LCS IV

Komentáře k této Příručce

Žádné komentáře