Calibration Procedure with Taber Abrasion Holder
Calibrate the instrument without the Taber Abrasion Holder.
Install the Taber Abrasion Holder in front of the sphere entrance.
Calibrate the instrument once more, but only dark- and bright calibration.
When you are asked to calibrate Clarity, please press the ”set-up“ button.
Perform measurements without a sample in Haze, Transmission and Clarity.
The result should be: H = 0.00, T = 100 and C = 100.
If the Taber Abrasion Holder is removed, calibrate the instrument again.
Note:
When working with the Taber Abrasion Holder, only measure Haze or Transmission.
Clarity measurements are not possible with the holder due to changed
optical conditons.