What to do if... BYK LC 4
48
No stable
measured value
Indication, l
Meter does not
react to keystroke
Sensor symbol
flashes
Message, St oFull
StoF ullStoF ull
StoF ull
6.1 Processor reset:
The processor reset is a simple action that can be helpful in the case of error
messages or error modes.
Cause Remedy
Sensor module contaminated Clean sensor module
Temperature not stable Adjust temperature if necessary
Conductivity too low Use suitable sensor module
Organic liquids Use suitable measuring system
Cause Remedy
Battery almost empty Exchange the battery (see
section 5.2)
Cause Remedy
Operating condition undefined or
EMC load unallowed
Perform a processor reset (see
section 6.1)
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
All memory locations are full Clear the memory (see section
4.9.3)
1 Press <1> and <8> at the same time and release them again.
Stored data and calibration values are retained.