15 Technical Data
Fe-Probe Measurements on iron and steel substrate
NFe-Probe Measurements on non-magnetic metal substrates, such as
Aluminum, Zinc, Copper, Brass, stainless steel
Range see specification of probes
Resolution 0.1 !m 0 to 99.9 !m 0.01 mil 0 to 9.99
mil
(standard setting) 1 !m 100 to 999 !m 0.1 mil 10 to 100
mil
0.01 mm " 1.00 mm 1 mil > 100 mil
Accuracy (factory calibration on zero reference plates supplied with the gauge)
± (1 !m + 2%*) 0 to 2000 !m
± (0.04 mil + 2%*) 0 to 80 mil
± 3.5%* > 2000 !m (80 mil)
(* of reading)
With the one-point or two-point calibration the accuracy can still be
improved using the optionally available test shims.
Minimum Fe-Probe 10 x 10 mm# (0.4" x 0.4")
Area NFe-Probe 6 x 6 mm# (0.24" x 0.24")
Minimum convex 5 mm (0.2")
Curvature concave 30 mm (1.2")
Minimum Fe-Probe 0.2 mm (8 mil)
Substrate Thickness NFe-Probe 0.05 mm (2 mil)
Memory Capacity Basic
Premium
Readings 100 up to 13,000 (max. 2000 readings per batch)
Batches 1 999
Calibrations 1 100
Interface wireless 2.4 GHz, range max. 10 m (30ft) in undisturbed environment
Temperature Range
Storage -10°C to 60°C (14°F to 140°F)
Operation 0°C to 50°C (32°F to 122°F)
Power Supply 2 x batteries (AA) 1.5V Alkali, or 2 x rechargeable batteries (AA) 1.2V
Dimensions 124 mm x 67 mm x 33 mm ( 4.9" x 2.6" x 1.3")
Weight ca. 120 g (4.3 oz) gauge with batteries and probe
byko-test 8500
Instruction Manual
15
Zobrazit stránku 0
1 2 3 4 5 6 7 8

Shrnutí obsahu

Strany 1 - Instruction Manual

15 Technical Data Fe-Probe Measurements on iron and steel substrate NFe-Probe Measurements on non-magnetic metal substrates, such as Aluminum, Z

Strany 2

MAIN MENU Measuring ModeData MemorySelect Substrate Fe / NFeNFeC ombined MeasurementMeasurement Type Single MeasurementC ontinuous ModeUpper LimitLo

Strany 3

5 Setting of Zero Reference When using the probe for the first time, after a battery exchange, with varying applications, or after a long period the

Strany 4 - 2 Change of Probes

7.3.4.2.2 Delete a Reading 10 7.3.4.3 Information About a Batch 11 7.3.4.4 Limits 11 7.3.4.5 Rename a Batch 11 7.3.4.6 Delete a Batch 11 7.4

Strany 5

7.3 Memory Mode With this option the memory mode is switched on or off. Also configuration and handling of the batches is controlled. After pressing k

Strany 6

7.1.1.4 Selection of User-Specific Calibration On this menu level all specific calibrations are listet below the option "New Calibration".

Strany 7

13 Symbols used in the Display ∞ Fe measurement on steel or iron substrate ∞ NFe measurement on non-magnetic substrate ∞ Error handling error ∞

Strany 8

7.5.1.2 Resolution The resolution of the readings can be adapted to the application. E.g. on surfaces with high roughness the resolution should be red